Test range 5mm to 20,000mm (0.25-800in) at steel velocity. Variable in 1,2,5, sequence, or continuously in 1mm (0.05in) increments
Reference Waveform This menu displays a waveform from one of the A-Log stores as a reference or fingerprint in a Colour
different from the active display highlighting differences from the reference.
DAC DAC curves may be entered and digitally drawn on the display. Reference, -6dB, -12dB,-14dB curves may be selected. DAC curve selected acts as gate for alarm outputs and height measurement in DAC +dB. DAC
parameters stored with Panel Memory. Curves comply with ASME and European Standards.
TCG Time Corrected Gain curves may be entered and digitally drawn on the display. The dynamic range of the T.C.G. is 40dB. T.C.G. parameters are stored in panel, A-Log and memory.
Auto-Cal Provides automatic calibration from two echoes
Notes Alphanumeric-labelling for panel and A-Log allows the user to enter Notes for storage with A-Scans.
Special Functions Display freeze for capturing current A-Scan image. Peak
memory for echo-dynamic pattern determination in accordance with BS3923. Help key for instant operator
guidance on using the Autoscan2400.
Help For instant operator guidance on using the Autoscan2400
Waveform Smoothing Gives a smooth signal envelope similar to the video filtering in analogue equipment.
Outputs Full bi-directional serial interface to transfer parameters, thickness readings and waveform memories. Composite video, full PAL or NTSC compatibility. Two analogue proportional programmable output signals (front panel) for connection to external equipment, such as a light alarm device. RF (80mVpp) and trigger outputs.
Power Lithium-Ion sealed battery pack. Typically 8 hours operation on colour display model at medium display brightness, more on LCD.
Mains power (220V) for continuous operation.
Charger Universal mains input 85-260 volts AC.
Probe Sockets LEMO 01 (factory option) or BNC.
Environment IP67 Keypad Membrane, with lock key function
(except for gain control).
Temperature Operating: -10 to +55°C (14 to 131°F)
Survivable: -20 to +70°C (-4 to 158°F)
Storage: -40 to +75°C (-40 to 167°F)
Size 255 x 145 x 145mm (10.0 x 5.7 x 5.7in)
Weight 2.5Kg (5.5 lbs.) with Li-ION cells.
Velocity range 1,000m/s to 9,999m/s continuously variable
Probe Zero 0 - 999.99 s continuously variable
Delay Calibrated delay from 0mm to l0,000mm in 0.05mm steps at steel velocity (0-400in. in 0.002in steps)
Gain 0 to 110dB. Adjustable in 0.5, 2, 6, 14 and 20dB steps. Direct access
Test Modes  
Pulser Square-wave pulse, 185V peak amp, 40ns-250ns linked to frequency
P.R.F. Selectable between 35, 63, 150, 250, 500, & 1,000Hz Update Rate 60Hz (NTSC Mode); 50Hz (PAL Mode)
Rectification Full wave, positive or negative half wave and unrectified rf Frequency Range Four (4) narrow bands centred at 1 MHz, 2MHz, 5MHz
and 10MHz. Wide band 1.5 to 15MHz (-6dB)
System Linearity Vertical ±1% Full Screen Height (FSH). Amplifier
accuracy ±0.1dB. Horizontal ±0.4% Full Screen Width
(FSW)
Reject 50% suppressive reject. LED warning light when selected
Units Metric (mm), inches (in) or microseconds. Selected from menu
Display High brightness Colour TFT LCD panel. Eight colour
scheme options. Brightness variable up to 350cd/m2.LCD: monochrome LCD, with switchable backlight. A-Scan area 255 x 200 pixels. Total display area 102.7 x 77.0mm, 320 x 234 pixels
Gate Monitor and Alarms Two fully independent gates for echo
monitoring and thickness measurement. Start and width
adjustable over full range of unit, amplitude variable from
0 to 100% FSH. Bar presentation. Positive or negative
triggering for each gate with visual and audible alarms.
Operator defined alarm thresholds (upper and lower limits of acoustic speed) in velocity measurement mode.
Gate expansion Expands range to width of Gate 1
Measurement Modes There are 6 modes: Signal monitor, Depth and
amplitude of first signal in gate, T-Min mode for holding
minimum thickness reading, Echo-to-echo distance
measurement, Trigonometric display of beam path,
surface distance and depth of indication Gate to Gate.
Resolution to 0.01mm (0.001in) for distance measurement, or 1% FSH for amplitude measurement. Large display of measurement at top of A-Scan display. Measurement mode selectable between peak and flank. All measurement functions available in unrectified ode.
Velocity Measurement Direct reading on screen of the acoustic velocity for
a given thickness.
A-Scan Memory Maximum of 100 waveforms stored with complete panel
settings. Waveforms may be recalled on display, printed or
transferred via RS232 serial interface
Panel Memories Twenty stores for retaining calibrations
Thickness Logging Storage for 2,000 thickness readings configured into
Block/Location/Number. Calibration settings stored with
each Block. Maximum number of Blocks is 14. Unlimited
Location/Number values, maximum combination of 2,000
readings. Readings may be reviewed, edited and printed
as required.

 
Get a competitive quotation
Get a competitive quotation
(price) on this product from
Sonatest or one of our many
authorized distributors.


Request more information about this product. This may be technical queries, applications that this product is used in or any other information you require related to this product.
Request more information about
this product. This may be
technical queries, applications
that this product is used in or
any other information you
require related to this product.


View a brochure on this product.
View a brochure on this product.

News
News Feed Bookmark and Share
Copyright 2010 Sonatest Limited | Shipping and Returns | Disclaimer | Privacy and Data Protection Policy | Site Map