Ultrasonic Specifications Measurement Modes:

Test Range: 0 - 1mm (0.05in) up to 0 - 20000 mm (800 in.) at steel velocity. Variable in 1,2,5 sequence or continuously in 1mm (0.05in) increments.

Velocity: 256 to 16000m/s continuously variable.

Probe Zero: 0 to 999.999 μs, continuously variable.

Delay: Calibrated delay from 0 -20000mm in 0.05 mm steps at steel velocity (0-400in. in 0.002 in. steps).

Gain: 0 to 110dB. Adjustable in 0.1, 0.5, 1, 2, 6, 10, 14 and 20dB steps.

Test Modes: Pulse echo and transmit/receive.

Pulser: 100V - 450V square wave pulser. Pulse width from Spike to 2500ns duration - rise/fall times <10ns into 50 ohms at 200V: Width adjustable in 2% of nominal width, minimum 1ns maximum 40ns.

ActiveEdge™ Unique active pulse control for enhanced near surface resolution and signal response.

P.R.F: Selectable 15 to 5000Hz in 5Hz steps.

Update Rate: 60Hz (NTSC Mode); 50Hz (PAL Mode).

Rectification: Full wave, positive or negative half wave and unrectified rf.

Frequency Range: 7 narrow bands centred at 0.25Mhz, 0.5MHz, 1MHz, 2.25MHz, 5MHz, 10MHz and 15MHz. Broad band at 2 MHz to 22MHz (-6dB) and 1MHz to 35 MHz (-20dB). Special Dryscan mode allowing the resonant frequency of the receiver stage of the Dryscan 410 to be adjusted between 150kHz and 1.42MHz in 10 kHz steps.

System Linearity: Vertical = 1% Full Screen Height (FSH). Amplifier Accuracy ± 0.1dB. Horizontal ±0.4% Full Screen Width (FSW).

Units: Metric (mm), inch (in) or time (μs).

 

Display:

 

Display: Colour Transflective TFT: Display area 111.4 x 83.5 mm (4.39 x 3.29 in) 320 x 240 pixels. A-Scan Area 255 x 200 pixels (315 x 200 expanded), 8 colour options and variable brightness.

Gate Monitor: Two fully independent gates for echo monitoring and thickness measurement. Start and width adjustable over full range of unit, amplitude variable from 0 to 100% FSH. Bar presentation. Positive or negative triggering for each gate with audible and visual alarms.

Measurement Modes

Mode 1:Signal Monitor

Mode 2:Depth and amplitude of first signal in gate.

Mode 3: Echo-to-Echo distance measurement. (single gate)

Resolution: To 0.01mm (0.001in) for distance measurement or 1% FSH for amplitude measurement. Large display of measurement at top of A-Scan display. Measurement mode selectable between peak and flank.

A-Scan Memory: Maximum of 800 waveforms can be printed or transferred to a PC using optional SDMS software.

Panel Memory: 100 stores for retaining calibrations.

 

DAC: DAC defined by up to 10 points and digitally drawn on screen. DAC curves meet requirements of EN 1714, JIS and ASME standards, selectable between -2, -6, -10, -12 and -14dB. Amplitude read out selectable between % DAC, % FSH or relative dB.

Auto-Cal: Provides automatic calibration from two echoes.

Clock: Sets time and date.

Reference Waveform: This menu displays a waveform from one of the A-log stores as a reference or fingerprint display in a colour different from the active display highlighting differences from the reference.

Notes: Alphanumeric labelling for panel and A-log allows the user to enter Notes for storage with panel settings and A-scans

Display Freeze: For capturing the current A-scan image.

Active Peak Memory: Envelope waveform for echo dynamic pattern determination, with simultaneous live A-scan display.

Keylock: Prevents accidental alteration of parameters.

Help Key: For instant operator guidance on using the Dryscan.

Language Support: Supports multiple languages. User selectable between English, German, Spanish, French, Dutch. Others available on request.

Waveform Smoothing: Gives a smooth signal envelope, simulating analogue equipment.

Outputs: Full bi-directional serial interface to transfer parameters, thickness readings and waveform memories. Composite video, PAL or NTSC compatibility. Analogue proportional outputs programmable to distance or amplitude of signal in the gate. Transmitter sync output.

Front USB: For connection to printers, keyboards and PC.

Printers: Supports Hp DeskJet, Epson.

Power: Lithium Ion battery pack 14.4V, 5.0 ampere hours, gives up to 16 hours duration from a fully charged pack. Indication of low battery status. Recharge time 3-4 hrs. Mains pack option.

Charger: 100 - 240 VAC, 50-60 Hz.

Transducer Sockets: BNC or LEMO (factory option)

Environmental: Case sealed to IP67

Temperature: Operating -10°C to +55°C (14 to 131°F). -20°C to +70°C. (-4 to 158°F) survivable. Storage: -40° to +75°C. (-40 to = 167°F)

 

Physical:

Size: 255 x 145 x 145mm (10.0 x 5.7 x 5.7 in)

Weight: 2.5kg (5.5lbs) with Li-Ion cells.

 

Accessories:

 

Standard Kit Includes Dryscan 410D Li-ion Battery & Battery Charger Fabric Carry Bag, Instruction Manual (EN12668)

 

Certification:

 

Calibration Certificate

 
Get a competitive quotation
Get a competitive quotation
(price) on this product from
Sonatest or one of our many
authorized distributors.


Request more information about this product. This may be technical queries, applications that this product is used in or any other information you require related to this product.
Request more information about
this product. This may be
technical queries, applications
that this product is used in or
any other information you
require related to this product.


View a brochure on this product.
View a brochure on this product.

News
News Feed Bookmark and Share
Copyright 2010 Sonatest Limited | Shipping and Returns | Disclaimer | Privacy and Data Protection Policy | Site Map