ZETEC MS 5800
|
 |
 |
|
| Power requirements: |
Voltage: 100 - 240 VAC,
±10%
Fuse: 250 V, 2 A, slow-blow
Voltage: 100 - 240 VAC, ±10%
Maximum input current: 1.5 A |
| Environmental conditions: |
Operating temperature:
0°C to 45° C (32°F to 113°F)
Storage temperature: -20°C to 60°C (-4°F
to 140°F)
Relative humidity: 95%, no condensation
Safety: EN61010-1 |
| MS 5800 housing: |
Size (W x H x D): 465.0
mm x 380.0 mm x 255.0 mm (18.3 x 15.0 x
10.0 inches)
Net weight: 12.8 kg (28.2 lb)
Protection: Static proof case |
| Ethernet cable: |
Type: Category 5, shielded
Maximum length: 100 m (328 ft) |
| Encoders: |
Number of encoder axes:
2
Encoder type: Quadrature or Clk/Direction
Quadrature resolution: 4 counts/cycle
Frequency adjustment resolution: <0.1%
Input type: TTL
Counter size: 32 bits (from -2,147,483,648
to +2,147,483,648 counts)
Maximum frequency input: 200 kHz
Immunity to noise: Programmable filter from
8 Hz to 250 kHz
Sample rate control: Programmable axis selection
and resolution
Input protection: Over voltage protection
and ±15 V continuous |
| Auxiliary Inputs: |
Type: Index, reset encoder,
external clock, enable acquisition, and
spare input
Index input: Used for C-scan index or rotation
synchronization
Preset encoder: Used to preset encoder counter;
programmable selection of preset counter.
External clock: Used to control sample rate
by external signal
Enable acquisition: Used to start/stop acquisition
Input type: TTL
Input protection: Overvoltage protection
and ±15 V continuous |
| Auxiliary Outputs: |
Type: External clock, relay
outputs, and spare output
External clock: Used to synchronize the
sample rate of other equipment with the
MS 5800
Relay outputs: Spare outputs: Defined by
the synchronization mode |
| Alarms: |
Type of data: Eddy current,
remote field, magnetic flux leakage, or
ultrasound raw data
Setting of alarms: Programmable
Maximum response time: Smaller than the
sampling period
Number of alarm outputs: 8 for electromagnetic
techniques / 3 for ultrasound techniques |
| Options: |
Eddy Current (ECT), Remote
Field (RFT), Magnetic Flux Leakage (MFL),
and Ultrasound (UT) |
|
|
|
|